SKU
HI3512-02

Dual-Input pH/ORP/ISE/EC/Resistivity/TDS/NaCl/Temperature Bench Meter

550,00 € 4.143,98 kn

Najniža cijena u zadnjih 30 dana: 550,00 EUR (4.143,98 kn)

Na lageru

HI3512 is a dual-input professional pH/ORP/ISE/EC/Resistivity/TDS/NaCl/Temperature bench meter with a graphic LCD designed to provide high accuracy and ease of use both in the laboratory as well as in harsh industrial conditions. This instruments is a dual-input model measuring pH, ORP, or ISE and temperature on channel 1 and EC, Resistivity, TDS, and Salinity (utilizing an EC probe) on channel 2. A Relative mV feature is also available.

This meter features Hanna’s exclusive Calibration Check™, a diagnostics system that ensures accurate readings every time. Calibration Check system eliminates erroneous readings due to dirty or faulty pH electrodes or contaminated buffer solution by alerting users of potential problems during the calibration process.
Throughout the calibration process, users are guided step-by-step by the on-screen tutorial. After calibration, a probe condition indicator is displayed on the LCD informing the user of the overall electrode status.
A variety of interactive user support is available before, during and after measurement. Onscreen tutorials guide users through set-up, calibration and measurement while context sensitive help of any screen is available at a push of a button. The HELP screen accessed by a dedicated HELP button, includes language specific assistance for menu parameters, calibration, log, contact information and accessories for your instrument.
Main Features:
• Dual-input channel
- pH/mV/ISE and temperature measurements (Channel 1)
- EC/TDS, NaCl/Resistivity and temperature measurements (Channel 2)
• Up to 5 point pH calibration with 7 standard buffers and 2 custom buffers to choose from
• Up to two points EC calibration with 7 memorized standards for calibration to choose from
• Messages on the graphic LCD for an easy and accurate calibration
• Diagnostic features to alert the user when the electrode needs cleaning
• Relative mV measurements
• Log on demand up to 400 samples
• Log interval with log on stability feature up to 600 records
• Auto Hold feature, to freeze first stable reading on the LCD
• GLP, last calibration data for pH, Rel mV, ISE, EC, or NaCl
• PC interface

SKU

HI3512

Product Name

Dual-Input pH/ORP/ISE/EC/Resistivity/TDS/NaCl/Temperature Bench Meter

pH Range

-2.0 to 20.0 pH / -2.00 to 20.00 pH / 2.000 to 20.000 pH

pH Resolution

0.1 pH / 0.01 pH / 0.001 pH

pH Accuracy

±0.01 pH / ±0.002 pH

pH Calibration

Up to 5 points / 7 Standard Buffers (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) / 2 Custom buffers

ORP Range

±2000.0 mV

ORP Resolution

0.1 mV

ORP Accuracy

±0.2 mV

ISE Range

1.00x10-7 to 9.99x1010 conc.

ISE Resolution

3 digits 0.01, 0.1, 1, 10 conc.

ISE Accuracy

±0.5% of reading (monovalent ions) / ±1% of reading (divalent ions)

ISE Calibration

Up to 5 points / 6 standards (0.1, 1, 10, 100, 1000, 10000 ppm)

Temperature Channel 1 Range

-20.0 to 120.0 °C (-4.0 to 248.0 °F)

Temperature Channel 1 Resolution

0.1 °C (0.1 °F)

Temperature Channel 1 Accuracy

±0.2 °C (±0.4 °F) (excluding probe error)

Rel mV offset range

±2000 mV

Slope calibration

From 80 to 110%

pH Temperature Compensation

Manual / Automatic

Electrode

Hanna BNC connection pH / ORP / ISE electrodes

Temperature Probe

RCA connection / Recommended option

EC Range

0 to 400 mS/cm (shows values up to 1000 mS/cm) / Actual conductivity 1000 mS/cm
0.001 to 9.999 μS/cm
10.00 to 99.99 μS/cm
100.0 to 999.9 μS/cm
1.000 to 9.999 mS/cm
10.00 to 99.99 mS/cm
100.0 to 999.9 mS/cm
1000 mS/cm (autoranging)

EC Resolution

0.001 μS/cm
0.01 μS/cm
0.1 μS/cm
0.001 mS/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm

EC Accuracy

±1% of reading (±0.01 µS/cm or 1 digit, whichever is greater) excluding probe error

EC Calibration

Up to 2 points / 7 Standards (0.00 µS/cm, 1.413 mS/cm, 5.00 mS/cm, 12.88 mS/cm, 111.8 mS/cm)

Resistivity Resolution

1.0 to 99.9 ohms
100 to 999 ohms
1.00 to 9.99 Kohms
10.0 to 99.9 Kohms
100 to 999 Kohms
1.00 to 9.99 Mohms
10.0 to 100.0 Mohms (auto-ranging)

Resistivity Accuracy

0.1 ohm
1 ohm
0.01 Kohms
0.1 Kohms
1 Kohms
0.01 Mohms
0.1 Mohms

Resistivity Calibration

±1% of reading (±10 ohms or 1 digit, whichever is greater) excluding probe error

TDS Range

0.000 to 9.999 ppm
10.00 to 99.99 ppm
100.0 to 999.9 ppm
1.000 to 9.999 g/L
10.00 to 99.99 g/L
100.0 to 400.0 g/L (autoranging)

TDS Resolution

0.001 ppm, 0.01 ppm, 0.1 ppm, / 0.001 g/L, 0.01 g/L, 0.1 g/L

TDS Accuracy

±1% of reading (±0.05 ppm or 1 digit, whichever is greater) excluding probe error

TDS Factor

0.40 to 1.00

Salinity Range

% NaCL: 0.0 to 400.0 %

Salinity Resolution

0.1%

Salinity Accuracy

±1% of reading (excluding probe error)

Temperature Channel 2 Range

-20.0 to 120.0 °C (-4.0 to 248.0 °F)

Temperature Channel 2 Resolution

0.1 °C (0.1 °F)

Temperature Channel 2 Accuracy

±0.2 °C (±0.4 °F) (excluding probe error)

NaCl Calibration

Max. 1 point only (with HI7073 standard)

EC probe

DIN connection / Recommended option

Temperature Source

Automatic / Manual

EC Temperature Compensation

NoTC, MTC, ATC

Reference temperature

15, 20, 25 °C

Log on demand

400 samples

Lot logging

5, 10, 30 seconds / 1, 2, 5, 10, 15, 30, 60, 120, 180 minutes, AutoEnd (maximum 600 samples)

Environment

0 to 50 °C (32 to 122 °F) / max. RH 55% non-condensing

Power Supply

12 Vdc power adapter

PC Interface

opto-isolated USB

Dimensions

235 x 207 x 110 mm (9.2 x 8.14 x 4.33 in.)

Weight

1.8 Kg (4.1 lbs)

Sve cijene su izražene bez PDV-a i bez troškova dostave, osim ako nije drugačije navedeno. HANNA Instruments d.o.o. — Jure Kaštelana 19 — 10000 Zagreb (Hrvatska) — OIB 57503101510 ©2024 Sva prava pridržana